12. At the CALIBRATION menu display, use
the Menu up-down keys to highlight
START CAL, then press the Select key.
13. At the PRECISION RETURN LOSS
CALIBRATION menu display prompt,
connect the Offset SWR Autotester to
Input A, if you have not done so yet.
14. Connect the precision air line to the Off
-
set SWR Autotester test port. Position
the air line pointing vertically upward.
Downward or horizontal positions make
connector pin alignment difficult.
NOTE
Ensure that the beadless end of the precision
airline is at the measurement connection
point.
15. Press the
Select key when ready.
16. At the PRECISION RETURN LOSS
CALIBRATION menu prompt, connect
the Open to the beadless end of the air
-
line. Press the
Select key to start the cali
-
bration.
17. Verify that the display resembles that
shown in Figure 4, page 14.
CAUTION
During both calibration and mea-
surement, be sure to properly
align the beadless connector of the
airline. When the connectors are
mis-aligned, a spike will usually
be visible on the display.
18. At the next menu prompt, remove the
Open and connect the Short to the
beadless end of the airline. Press the
Se
-
lect
key to start the calibration process.
19. At the next menu prompt, remove the
Short and connect the 50 Ohm Termina
-
tion to the beadless end of the air line.
Press the
Select key to start the calibra
-
tion process.
20. When the calibration is complete, remove
the 50 Ohm Termination.
21. Connect the SM/PL termination to the
beadless end of the air line and press the
Select key to begin the measurement.
22. Observe that the waveform displayed re
-
sembles that shown in Figure 5.
23. Press the
Cursor On/Off key to On.
24. Observe the Cursor menu readout. The
minimum return loss reading for the
SM/PL termination should be 42 dB.
15. INSTACAL MODULE VERIFICATION
This test verifies the performance of the Anritsu Site
Master InstaCal Calibration Module. The InstaCal
Module, part number ICN50, is an optional accessory
for the S114C and S332C.
NOTE
Full verification of the InstaCal module over
its entire frequency range (2 - 4000 MHz) re
-
quires both an S114C and S332C Site Master.
However, limited verification is possible with
either instrument for the frequency range of
that instrument only. The frequency range of
the S114C is 2 to 1600 MHz. The frequency
range of the S332C is 25 to 4000 MHz.
14 Site Master S114C/S332C MM
1 0 . 0 d B / D I V O F F S E T1 : P R E C S N R L ( A )
2 : O F F
+ 0 . 0 d B
0 . 4 G H z / D I V
S T A R T : 0 . 0 1 0 0 G H z
4 0 1 p t s
S T O P : 4 . 0 0 0 0 G H z
L E V E L + 7 . 0 d B m
5 4 1 4 7 A
P R E C I S I O N
R E T U R N L O S S
C A L I B R A T I O N
C O N N E C T O P E N
T O A I R L I N E
P R E S S S E L E C T
W H E N R E A D Y
Figure 4. Example of a Good Connection
1 0 . 0 d B / D I V O F F S E T1 : P R E C S N R L ( A )
2 : O F F
+ 0 . 0 d B
0 . 4 G H z / D I V
S T A R T : 0 . 0 1 0 0 G H z
4 0 1 p t s
S T O P : 4 . 0 0 0 0 G H z
L E V E L + 7 . 0 d B m
5 4 1 4 7 A
C U R S O R
1 : - 4 6 . 0 2 d B
P R E S S S E L E C T
F O R
C U R S O R M E N U
2 . 4 0 5 0 G H z
Figure 5. Direct Readout of the Precision Return Loss