Shrnutí obsahu

Strany 1 - Cable Analyzer

Site MasterModel S251CAntenna andCable AnalyzerUser's GuideHand-Held Tester for Transmission Linesand other RF ComponentsColor Front Cover only P

Strany 2

·Armored Test Port Extension Cable, 1.5 meter, N (m) toN (m), Part No. 15NN50-1.5C·Armored Test Port Extension Cable, 3.0 meter, N (m) toN (m), Part N

Strany 3 - TRADEMARK ACKNOWLEDGMENTS

1-4Chapter 1 General InformationSpecifications are valid when the unit is calibrated at ambient temperature after a 5 minutewarmup.DescriptionValueSit

Strany 4

1-5Chapter 1 General InformationCable Insertion Loss:RangeResolution0.00 to 54.00 dB0.01 dBBias Teeinput voltage:output current:12.5 to 15.0 Vdc270 ma

Strany 5 - Table of Contents

Preventive MaintenanceSite Master preventive maintenance consists of cleaning the unit and inspecting and clean-ing the RF connectors on the instrumen

Strany 6

Annual VerificationAnritsu recommends an annual calibration and performance verification of the Site Masterand the OSL calibration components by local

Strany 7 - Appendix B - Windowing

1-8Chapter 1 General InformationUNITED STATESANRITSU COMPANY685 Jarvis DriveMorgan Hill, CA 95037-2809Telephone: (408) 776-8300FAX: 408-776-1744ANRITS

Strany 8 - General Information

Chapter 2Functions and OperationsIntroductionThis chapter provides a brief overview of the Site Master functions and operations, provid-ing the user w

Strany 9 - Optional Accessories

Front Panel OverviewThe Site Master menu-driven user interface is easy to use and requires little training. Hardkeys on the front panel are used to in

Strany 10 - Performance Specifications

Function Hard KeysMODE Opens the mode selection box (below). Use the Up/Down arrow key to select amode. Press theENTER key to implement.FREQ/DIST Disp

Strany 11 - Chapter 1 General Information

Keypad Hard KeysThis section contains an alphabetical listing of the Site Master front panel keypad controlsalong with a brief description of each. Wh

Strany 13 - ESD Precautions

The following keypad hard key functions are printed in blue on the keypad keys.Turns the liquid crystal display (LCD) back-lighting ON or OFF.(Leaving

Strany 14 - Annual Verification

Soft KeysEach function key opens a set of soft key selections. Each of the soft keys has a correspond-ing soft key label area on the display. The labe

Strany 15

Chapter 2 Functions and Operations2-7MODE=DTF:SOFTKEYS:BOTTOMTOPFREQ/DISTAMPLITUDED2DTF AIDMORED1LOSSCABLEWINDOWBACKPROPVEL130259517PAGE UPSELECTTRACE

Strany 16 - Functions and Operations

FREQ/DIST Displays the frequency and distance menu depending on the measurement mode.FrequencyMenuProvides for setting sweep frequency end points when

Strany 17 - Front Panel Overview

AMPLITUDE Displays the amplitude or scale menu depending on the measurement mode.AmplitudeMenuProvides for changing the display scale. Selected values

Strany 18 - Function Hard Keys

Power Monitor Menu (Option 5)Selecting POWER MONITOR from the Mode menu causes the soft keys, described below,to be displayed and the corresponding va

Strany 19 - Keypad Hard Keys

RF Source MenuRF Source mode provides an RF Out source that can be used, for example, in the alignmentof dish antennas. SelectingRF SOURCE from the Mo

Strany 20

LIMITPressing LIMIT on the data keypad activates a menu of limit related functions.Use the corresponding softkey to select the desired limit function.

Strany 21 - Soft Keys

qEDIT — Opens the selected marker parameter for data entry. PressENTER when data entry is complete or ESCAPE to restore the previousvalue.qDELTA (Mx-M

Strany 22 - AMPLITUDE

SEGMENT 3SEGMENT 4SEGMENT 5BACKqLIMIT BEEP — Turns the audible limit beep indicator on or off.SYS Displays the System menu softkey selections.qOPTIONS

Strany 23

WARRANTYThe Anritsu product(s) listed on the title page is (are) warranted against defects inmaterials and workmanship for one year from the date of s

Strany 24

SymbolsTable 2-1 provides a listing of the symbols used as condition indicators on the LCD display.Self TestAt turn-on, the Site Master runs through a

Strany 25 - Power Monitor Menu (Option 5)

Error CodesSelf Test ErrorsA listing of Self Test Error messages is given in Table 2-2.2-16Chapter 2 Functions and OperationsError MessageDescriptionB

Strany 26 - RF Source Menu

Range ErrorsA listing of Range Error messages is given in Table 2-3.2-17Chapter 2 Functions and OperationsError MessageDescriptionRANGEERROR:F1 > F

Strany 27

Battery InformationCharging a New BatteryThe NiMH battery supplied with the Site Master has already completed three charge anddischarge cycles at the

Strany 28

The NiMH battery will last longer and perform better if allowed to completely dischargebefore recharging. For maximum battery life, it is recommended

Strany 29

The battery can be charged and discharged 300 to 500 times, but it will eventually wear out.The battery may need to be replaced when the operating tim

Strany 30 - Self Test

Chapter 3Getting StartedIntroductionThis chapter provides a brief overview of the Anritsu Site Master. The intent of this chapteris to provide the use

Strany 31 - Error Codes

Select the Frequency or DistanceRegardless of the calibration method used, the frequency range for the desired measure-ments must be set before calibr

Strany 32 - Range Errors

Calibration Procedure - Single PortCheck that the “CAL OFF” message is displayed in the upper left corner of the display.This indicates that the Site

Strany 33 - Battery Information

Calibration Procedure - Dual PortCheck that the “CAL OFF” message is displayed in the upper left corner of the display.This indicates that the Site Ma

Strany 35 - Important Battery Information

Setting the ScaleAuto ScaleThe Site Master can automatically set the scales to the minimum and maximum values ofthe measurement on the y-axis of the d

Strany 36 - Getting Started

Step 1. Press the SAVE DISPLAY key to activate the alphanumeric menu for trace stor-age.Step 2. Use the soft keys to enter a label for the saved trace

Strany 37 - Calibration

Set the Distance and Cable TypeIn Distance-To-Fault (DTF) mode, the length of the transmission line (distance) and cabletype are selected. The cable t

Strany 38 - Figure 3-2. Calibration Setup

PrintingPrinting is accomplished by selecting an available printer and pressing the print key as de-scribed below. Refer to the particular printer ope

Strany 39 - Site Master S251C

Step 4. Press the PRINTER soft key and select from the displayed menu of supportedprinters.Step 5. Press thePRINT key. (Figure 3-6).Printer Switch Set

Strany 40 - Save and Recall a Display

Using the Soft Carrying CaseThe soft carrying case has been designed such that the strap can be unsnapped to allow thecase to be easily oriented horiz

Strany 41 - Recalling a Display

Chapter 4Cable & AntennaAnalysisIntroductionThis chapter provides a description of cable and antenna measurements in one port and twoport modes. M

Strany 42 - Changing the Display Language

The performance of a transmission feed line system may be affected by excessive signal re-flection and cable loss. Signal reflection occurs when the R

Strany 43 - Printing

Typical Line Sweep Test ProceduresThis section provides typical line sweep measurements used to analyze the performance ofa transmission feed line sys

Strany 44 - Printer Switch Settings

Figure 4-2 is an example of a typical system return loss trace:NOTE: The system sweep trace should appear at an approximate return loss of15 dB (±3 dB

Strany 45 - Using the Soft Carrying Case

Table of ContentsChapter 1 - General InformationIntroduction ...1-1Description...1-1Sta

Strany 46 - Analysis

Insertion Loss MeasurementThe transmission feed line insertion loss test verifies the signal attenuation level of the ca-ble system in reference to th

Strany 47

Step 14. Verify the measured insertion loss against the calculated insertion loss:Type Attenuation (dB/ft)´ Length (ft) = Loss (dB)First JumperLDF4-50

Strany 48 - MODE key

Procedure - Return Loss ModeStep 1. Press theMODE key.Step 2.Select FREQ-RETURN LOSS using the Up/Down arrow key and pressENTER.Step 3. Set the start

Strany 49

Distance-To-Fault (DTF) Transmission Line TestThe Distance-To-Fault transmission line test verifies the performance of the transmissionline assembly a

Strany 50 - Insertion Loss Measurement

Figure 4-5 shows an example of a typical DTF return loss trace.In the above example:qMarker M1 marks the first connector, the end of the Site Master p

Strany 51 - ´ Length (ft) = Loss (dB)

Antenna Subsystem Return Loss TestAntenna Subsystem return loss measurement verifies the performance of the transmit andreceive antennas. This measure

Strany 52

The following trace is an example of an antenna return loss trace.Calculate the threshold value and compare the recorded Lowest Return Loss to the cal

Strany 53

Two Port MeasurementsIntroductionA Tower Mounted Amplifier (TMA) can be used to amplify the received signal. There aredifferent types of TMA depending

Strany 54

TMA MeasurementsRequired EquipmentqSite Master Model S251CqOption 10A, Bias TeeqExternal Power Supply 40-115qTest Port Extension Cable, Anritsu 15NNF5

Strany 55 - MARKER key

Step 8.Press theSYS key, then OPTIONS, then BIAS TEE to turn on the bias tee volt-age. The gain is the level change before and after the bias tee volt

Strany 56

Saving a Setup ...3-5Recalling a Setup...3-5Save and Recall a Display ...

Strany 57 - Two Port Measurements

Device Under TestqTMA-DDProcedureStep 1. Connect the External Power Supply (40-115) to the Site Master.Step 2. Turn on the Site Master and press theMO

Strany 58 - TMA Measurements

Marker M1 indicates the receive band of the tower mounted amplifier. The level above the0 dB reference level is the gain of the TMA. Marker M2 indicat

Strany 59 - Gain/Insertion Loss

Device Under TestqTMA-DProcedureStep 1. Connect the External Power Supply (40-115) to the Site Master.Step 2. Turn on the Site Master and press theMOD

Strany 60 - RX/TX ANT

Marker M1 indicates the receive band of the tower mounted amplifier. The level above the0 dB reference level is the gain of the TMA. Marker M2 indicat

Strany 61 - Chapter 4 Measurements

Antenna to Antenna IsolationIntroductionMeasuring antenna-to-antenna isolation between adjacent systems is very important whendeploying multiple syste

Strany 62 - (BIAS TEE)

Antenna-to-Antenna Isolation MeasurementAntenna-to-Antenna isolation measurement is accomplished in Insertion Loss Mode. Inser-tion Loss Mode (+ 6 dBm

Strany 63

Step 5. Disconnect the Tx (Transmit) coax from System 1 and connect it to the RF Outport of the Site Master.Step 6. Disconnect the Rx (Receive) coax f

Strany 64 - Antenna to Antenna Isolation

Chapter 5Site Master Software ToolsIntroductionThis chapter provides a description of the Site Master Software Tools provided with theSite Master. Sit

Strany 65

InstallationTo install the Site Master Software Tools program:Step 1. Insert the Anritsu Site Master Software Tools disk in the CDROM drive.Step 2. Fr

Strany 66 - 4-21/4-22

Interface Cable InstallationCommunication between the Site Master and the PC is accomplished over a null modem se-rial cable provided with the Site Ma

Strany 67 - Site Master Software Tools

Appendix A - Reference DataCoaxial Cable Technical Data...A-1Appendix B - WindowingIntroduction ...

Strany 68 - Installation

Plot CaptureTo open the plot capture menus, select the capture icon on the button bar, or select the Cap-turedrop down menu from the menu bar.SelectCa

Strany 69 - Using Software Tools

MarkersMarkers M1 through M6 can be activated from the Plot Properties menu.The six active markers can be displayed on the plot to help identify fault

Strany 70 - Plot Properties

Step 3. Exit the Site Master Software Tools program and open the target application(Microsoft Word, etc.).Step 4. SelectEdit and then Paste. The file

Strany 71 - Saving Traces

Uploading a Cable ListAfter custom cable list has be created in Site Master Software Tools it can upload it to theSite Master. The serial communicatio

Strany 72 - Custom Cable List

Appendix AReference DataCoaxial Cable Technical DataTable A-1 provides a standard listing of common coaxial cables along with their RelativePropagatio

Strany 73 - Printing Formats

A-2Appendix A Reference DataManufacturer Cable TypeRelativePropagationVelocity (Vf)NominalAttenuationdB/m @1000 MHzNominalAttenuationdB/m @2000 MHzNom

Strany 74 - Reference Data

A-3/A-4Appendix A Reference DataManufacturer Cable TypeRelativePropagationVelocity (Vf)NominalAttenuationdB/m @1000 MHzNominalAttenuationdB/m @2000 MH

Strany 75 - Appendix A Reference Data

Appendix BWindowingIntroductionThe Distance sub-menu (page 2-8) provides for setting the cable loss and relative propaga-tion velocity of the coaxial

Strany 76

B-2Appendix B WindowingD i s t a n c e T o F a u l t51 01 52 0 2 53 0F e e t3 54 04 55 05 56 0- 5- 1 0- 1 5- 2 0- 2 5- 3 0- 3 5- 4 0- 4 5- 5 0R e

Strany 77 - Windowing

B-3/B-4Appendix B WindowingD i s t a n c e T o F a u l t51 01 52 0 2 53 0 3 5 4 04 55 05 5 6 0F e e t- 5- 1 0- 1 5- 2 0- 2 5- 3 0- 3 5- 4 0- 4 5-

Strany 78 - Appendix B Windowing

Chapter 1General InformationIntroductionThis chapter provides a description, performance specifications, optional accessories, pre-ventive maintenance

Strany 81

The following items are supplied with the basic hardware.·Soft Carrying Case·AC-DC Adapter·Automotive Cigarette Lighter 12 Volt DC Adapter,·CDROM disk