Site Master™Models S114C and S332CAntenna, Cable and Spectrum AnalyzersMAINTENANCE MANUAL490 JARVIS DRIVE · MORGAN HILL, CA 95037-2809P/N: 10580-00068

Strany 2

19. Press the FREQ/DIST key and the STARTsoft key.20. Enter4 and press the MHz soft key to setthe start frequency to 4 MHz.21. Press theSTOP soft key

Strany 3 - Table of Contents

9. Press the FREQ/DIST key and the CENTERsoft key.10. Enter1 and press the GHz soft key to setthe center frequency to 1GHz.NOTEThe Measured Occupied B

Strany 4

10. SPECTRUM ANALYZER MEASUREMENTACCURACY TESTThis test verifies the level accuracy of the Site Mas-ter Spectrum Analyzer.a. Equipment Required:·Anrit

Strany 5 - Site Master S114C/S332C MM 1

12. Press the SPAN soft key and enter .5,then press theMHz key to set the span to.5 MHz.13. Press theAMPLITUDE key.14. Press theREF LEVEL soft key, en

Strany 6 - 2 Site Master S114C/S332C MM

11. DISPLAYED AVERAGE NOISE LEVELThe following test can be used to verify the displayedaverage noise level of the Site Master Spectrum Ana-lyzer.a. Eq

Strany 7 - Site Master S114C/S332C MM 3

34. Press the STOP soft key.35. For the S114C, enter1600 then press theMHz soft key to set the stop frequency to1600 MHz.For the S332C, enter2000 then

Strany 8 - 4 Site Master S114C/S332C MM

16. Adjust the signal levels of both signalsources so that both signal peaks are at–20 dBm as indicated on the Site MasterSpectrum Analyzer display.17

Strany 9 - Site Master S114C/S332C MM 5

14. TERMINATION VERIFICATIONThis test verifies the accuracy of the Site MasterSM/PL termination using the precision return lossmode of the 541XXA Scal

Strany 10 - 6 Site Master S114C/S332C MM

12. At the CALIBRATION menu display, usethe Menu up-down keys to highlightSTART CAL, then press the Select key.13. At the PRECISION RETURN LOSSCALIBRA

Strany 11 - Site Master S114C/S332C MM 7

a. Equipment Required:·InstaCal Module, part number ICN50·20 dB offset, Anritsu SC5270·6 dB offset, Anritsu SC5237b. ProcedureNOTEIf performing full v

Strany 13 - Site Master S114C/S332C MM 9

16. BATTERY PACK REMOVAL ANDREPLACEMENTThis procedure provides instructions for removingand replacing the Site Master battery pack.NOTEProcedures in t

Strany 14 - 10 Site Master S114C/S332C MM

17. BATTERY INFORMATIONThe following information relates to the care andhandling of the Site Master battery, and NiMH bat-teries in general.·The Nicke

Strany 15 - Site Master S114C/S332C MM 11

18. FRONT PANEL ASSEMBLY REMOVALAND REPLACEMENTThis procedure provides instructions for removingand replacing the Site Master front panel assembly.Wit

Strany 16 - Site Master S331B

19. LCD ASSEMBLY REPLACEMENTThis procedure provides instructions for removingand replacing the Liquid Crystal Display (LCD) oncethe front panel assemb

Strany 17 - SWR Autotester

21. KEY PAD MEMBRANE REPLACEMENTThis procedure provides instructions for replacingthe key pad membrane.1. Remove the front panel assembly as directedi

Strany 18

23. REPLACEABLE PARTSReplaceable parts for the Site Master Models S114Cand S332C are listed below.Site Master S114C/S332C MM 21Part Number Description

Strany 19 - Site Master S114C/S332C MM 15

23.1. Anritsu Customer ServiceOnline support and updates to the product softwareor documentation, if any, can be downloaded from theAnritsu Web site a

Strany 22 - RAISED AREA

Table of Contents1. INTRODUCTION ...42. DESCRIPTION ...43. SITE MASTER FREQ

Strany 24 - 20 Site Master S114C/S332C MM

1. INTRODUCTIONThis manual provides maintenance instructions forthe Site Master Model S114C and S332C Antenna,Cable, and Spectrum Analyzers. It descri

Strany 25 - Site Master S114C/S332C MM 21

4. RETURN LOSS VERIFICATIONThe following test can be used to verify the accuracyof return loss measurements. Measurement calibra-tion of the Site Mast

Strany 26

15. Press the VBW soft key and then theMANUAL soft key. Use the Up/down arrowkey to select3 kHz. Press ENTER to setthe video bandwidth to 3 kHz.16. Pr

Strany 27

NOTEThe measured value is converted to dBc/Hz us-ing the following formula:()[]dBc Hz measureddBcRBWHz/log=- - 101At 10 kHz RBW,()10 401logRBWHz=,sodB

Strany 28

19. On the MG3692A Synthesized SignalSource, adjust the output level so thatthe M1 reading of Site Master SpectrumAnalyzer is –30 dBm at 40 MHz.20. On