4-4
Section 4 OPERATING INSTRUCTIONS
4.2.1 Logic modification
① The logic of the set pattern is changed in positive → nega-
tive → positive order each time the LOGIC key is pressed.
(The set logic is indicated by lighting of a lamp.)
4.2.2 Alternate pattern setting
① Select ALTN with the keys.
(DATA, Z.S, and PRBS are also selected with these keys.)
ALTN → DATA → Z.S. → PRBS
ALTN ← DATA ← Z.S. ← PRBS
② Pattern A and pattern B are selected with this key. Since
pattern A is set first, set this key so that the A lamp lights.
(It does not matter which pattern is set first.)
③ Set DATA LENGTH with the and keys.
This value is common to both patterns A and B.
Select the digit to be set with the keys.
Set DATA LENGTH with the keys.
Set value: 128 to 4,194,304 bits (128-bit steps)
④ Set the number of pattern A loops with the keys.
N
N
PATTERN
LOGIC PATTERN
PRBS / ZERO SUBST
PRBS MARK RATIO
2 - 1
2
POS
NEG
ALTNDATA Z.S PRBS 7 9 11 15 20 23 31 0/8 1/8 1/4 1/2
8/8 7/8 3/4 1/2
①
PATTERN
LOGIC PATTERN
PRBS / ZERO SUBST PRBS MARK RATIO
2 - 1
N
2
N
POS
NEG
ALTNDATA Z.S PRBS 7 9 11 15 20 23 31 0/8 1/8 1/4 1/2
8/8 7/8 3/4 1/2
ALTN
ERROR ADDITION(1×10 )
- N
SINGLE
AB
ON
4 5 6 7 8 9
①
②
④
DISPLAY
DATA LENGTH/ZERO SUBST LENGTH
DATA LENGTH
ZERO SUBST LENGTH
A / B LOOP TIME
PAGE/ PATTERN SYNC POSITION
③
DISPLAY
DATA LENGTH/ZERO SUBST LENGTH
DATA LENGTH
ZERO SUBST LENGTH
A / B LOOP TIME
PAGE/ PATTERN SYNC POSITION
Komentáře k této Příručce