Anritsu MG3642A Uživatelský manuál Strana 120

  • Stažení
  • Přidat do mých příruček
  • Tisk
  • Strana
    / 212
  • Tabulka s obsahem
  • KNIHY
  • Hodnocené. / 5. Na základě hodnocení zákazníků
Zobrazit stránku 119
SECTION 6 GPIB
6-14
6.3 Connecting the GPIB Cable
A GPIB cable must be connected for remote control by the GPIB. The MG3641A/MG3642A supports the GPIB
cable connector on the rear panel.
CAUTION
Turn the POWER switch off and unplug the power cord before connecting and disconnecting the GPIB cable.
This is because only the signal common line of the cord may rarely be cut before the others depending the cord
connection-disconnection method. In this case, if the power remains turned on, the AC leak voltage, etc. is
superimposed to the IC, and the circuit parts may be damaged.
A system using GPIB is restricted as follows. The device must be connected according to the following conditions:
Number of connectable devices 15
Total cable length 2 m × Number of devices (max. 20 m)
MG3641A/MG3642A rear panel
GPIB connector
GPIB cable
Zobrazit stránku 119
1 2 ... 115 116 117 118 119 120 121 122 123 124 125 ... 211 212

Komentáře k této Příručce

Žádné komentáře