TROUBLESHOOTING SYSTEM HAS ERROR CODE
373XXA MM 5-13
Item Description
SWP Sweep direction:
F = forward (S11, S21), R = reverse (S22, S12)
PWR System source power setting
SYS System frequency setting
A The calibration results for the Test A Channel PCB (A3): This is a
series of 16 values divided into groups of four. Each group should
have values that are approximately
1
4
that of those in the previous
group.
The first group data values should be approximately: 1.00 0.00
0.00 1.00. The values of the next group should be
1
4
that of those
values i.e., 0.25 0.00 0.00 0.25. The same applies for the values
in the remaining two groups.
B The calibration results for the Test B Channel PCB (A6). See de
-
scription for “A” above.
R The calibration results for the Reference A (or B) Channel PCB
(A4). See description for “A” above.
<OVL>
Indicates the value was too far out of range to be measurable.
PHSLCK Source Locking: L=Source Lock Mode, T=Tracking Mode, S=Set On
Mode
Table 5-3. Service Log Snap Shot Data Items